Shore Hardness Tester Requirements for US Purchasing Teams
Specify the required test method and scale, specimen-compatible instrument and fixtures, reporting, and calibration documentation; separate these requirements from recommended stands and optional workflow features. Bareiss identifies the product specification—not a preferred instrument—as the starting point for selection.[1][2]

- Required by the specification or quality procedure: Record the method, scale, specimen conditions, equipment configuration and documentation your controlling documents call for.
- Recommended for the intended use: Identify recommendations separately, including ZwickRoell’s laboratory test-stand recommendation.[1]
- Optional unless your workflow requires them: List digital displays, PC connections and automation separately from method and specimen-fit requirements.[1][2]
Use the same checklist for an existing setup: record its documented capabilities and unresolved mismatches before relying on it for incoming inspection.
Read the material specification before naming a scale
Start the equipment inquiry with the controlling product specification, required test standard and edition, scale or method, and hardness acceptance range. These are proposed purchasing fields, not universally prescribed certificate fields. Bareiss’s guide says selection should begin with the product specification and test method.[2]
Shore A generally addresses softer rubber and elastomers; Shore D addresses harder rubber and many hard plastics. These categories help orient selection, but readings on different Shore scales are not interchangeable. Bareiss also says Shore is not an equivalent substitute when the specification requires IRHD.[2]
ZwickRoell describes ASTM D2240 as covering a broad range of materials and durometer types, ISO 48-4 as focusing on vulcanized or thermoplastic rubber, and ISO 868 as addressing plastics and ebonite, particularly Shore A and D.[1] Check the applicable specification and standard edition rather than treating these manufacturer summaries as the controlling requirements.[1][2]
If the method or acceptance range is unclear, resolve it with the responsible quality or engineering function before finalizing the inquiry.
Match specimen geometry and test-point space
Attach a drawing, dimensions or a representative specimen description. Identify thickness, curvature, available test area, and whether the test location is flat, thin, small or irregular. Bareiss identifies these as selection factors, along with specialized fixture needs.[2]
Lay out the proposed indentations using the minimum distances between test points and from the specimen edge in the controlling method. Use that layout when assessing the available test area and fixture. The suppliers’ general method summaries help identify what to check. [1][2]
Keep the method decision separate from the fixture decision. Bareiss identifies Micro Shore for small or thin components and IRHD for specified or geometry-related applications; geometry alone does not authorize replacing the specified method.[2]
As a geometry-accommodation example, Germany-based Bareiss says its digi test II accepts interchangeable measuring devices for convex and concave specimens. This is not blanket confirmation that every curved part can be tested.[2] Request a proposed instrument-and-fixture configuration addressing the specimen conditions submitted, rather than accepting a stand alone as the answer to thickness or curvature concerns.
Resolve method-specific geometry exceptions against the applicable specification and standard.[1][2]
Separate stand requirements from optional features
Specify a stand when the controlling method or quality procedure calls for one. ZwickRoell recommends combining the instrument with a stand for laboratory testing, but also describes portable measurements under ASTM D2240, ISO 48-4 and ISO 868. Its laboratory recommendation is not a universal stand requirement.[1]
State whether use will be portable or benchtop, describe specimen support and positioning needs, and request an itemized instrument, stand and fixture configuration suited to those conditions.[1][2]
- Specimen fit: Evaluate the complete configuration against the submitted dimensions and test locations, including any specialized fixture needs.[2]
- Stand configuration: Separate a method- or procedure-driven requirement from the laboratory recommendation. Request the proposed stand’s loading specification and confirmation of instrument and fixture compatibility.
- Display and connectivity: ZwickRoell offers analog and digital instruments and PC-connection options. Choose these according to how operators need to read and transfer results.[1]
- Automation: ZwickRoell and Bareiss describe automated configurations. Assess them against test volume and acceptable operator involvement, separately from specimen compatibility.[1][2]
Mark workflow features as purchasing requirements only where the documented workflow needs them; keep the method and specimen-fit assessment separate.
Define the inspection record
Decide what the inspection record must contain before choosing how to capture it. ZwickRoell describes hardness notation as the numerical value, “Shore” and the method designation, with a nonstandard dwell time added as a fourth component—for example, 75 Shore A 10.[1]
Use the inquiry to define any additional internal reporting requirements, without presenting them as universally mandated standard fields:
- Specimen or lot identification.
- Test-location identification.
- Operator information.
- Results to retain and any required file format.
Take the dwell time, number of readings and rule for combining results from the controlling method. Use ZwickRoell’s comparison to identify where those settings differ across methods. [1]
Then assess whether manual recording meets those needs or electronic transfer is necessary. ZwickRoell’s PC-connection options should be evaluated against the required record, not treated as a reporting specification in themselves.[1]
Specify calibration documentation separately
Add a separate calibration-documentation line item: identify the instrument and applicable method or scale, then state the organization’s traceability and accreditation requirements. Replace “calibration certificate included” with a request the supplier can quote against.
ZwickRoell identifies ISO 21509 and ISO 48-9 in relation to instrument calibration.[1] Bareiss advertises ISO/IEC 17025-accredited calibration services for applicable hardness instruments.[2] A reference to a calibration standard and an offer of accredited calibration are distinct claims.
Request a sample certificate before ordering and compare it with internal quality requirements. Treat this as a purchasing safeguard, not a supplier-established approval process:
- Required documentation: Record what the quality system or customer specification calls for.
- Quoted documentation: Have the supplier identify the calibration service and documents offered for the specified instrument and scale.
- Unresolved differences: Identify and settle gaps before accepting the documentation package.
Ask the supplier to confirm certificate contents, traceability-chain documents, calibration intervals, due-date labels and as-found/as-left provisions against your controlling quality documents; specify routine verification and specimen-level records separately from accredited calibration.[1][2]
Keep check-block verification distinct from calibration
Include check blocks as a separate inquiry item if the routine-verification procedure calls for them, not as an automatically required accessory. Ask the supplier to identify the proposed block’s role and documentation, and compare these with your procedure’s acceptance values, tolerances, checking schedules, required records and failed-check rules.[1][2]
Before specifying a block, settle these fields in the internal verification procedure rather than assuming limits:
- Identification and documentation: Identify the block and supporting documents the procedure requires.
- Target values and tolerances: Define comparison values and acceptance limits.
- Frequency: State when verification is required.
- Records: Specify what the operator must document.
- Failed checks: Define the response to results outside the established limits.
Keep instrument calibration, routine verification and specimen test records as separate checklist entries. Check any proposed substitution of a block for calibration paperwork or calibration paperwork for the incoming-inspection record against your controlling quality documents.[1][2]
Prepare the equipment inquiry
Ask for an itemized response separating the base instrument, necessary stand or fixture, calibration documentation and optional accessories. The following is a recommended inquiry format, not a manufacturer’s package or a reproduced standard.
For each entry, record the requirement, proposed configuration or evidence, and status: required and addressed, optional, or unresolved.
- Application: Material or part, inspection purpose, controlling specification identifier and revision.
- Method and scale: Required method, scale and standard edition; proposed instrument capability.
- Hardness range: Specified range and acceptance criteria; proposed measuring capability.
- Specimen geometry: Thickness, curvature, available area and test locations; evidence the configuration accommodates them.
- Instrument, stand and fixture: Required use conditions; itemized proposed configuration and compatibility evidence.
- Reporting: Required result format and specimen-record fields; display, output capability and sample record.
- Calibration: Required certificate, traceability and supporting documentation; proposed service and sample documents.
- Routine checks: Procedure, blocks, limits, frequency and records specified internally; proposed provisions.
- Optional features: Data connections, automation or other preferences, itemized separately.
For an existing setup, enter documented capabilities and missing evidence instead of a supplier proposal. Resolve uncertainties about the method, specimen compatibility and documentation before treating either configuration as a match.
References
- ZwickRoell — Shore Hardness Testing and Durometers (zwickroell.com)
- Bareiss — Choosing the Best Hardness Tester (bareiss-testing.com)