Pressure Gauge Calibration: Compare Labs for Canadian Teams
Compare WIKA and Duncan on capability, accreditation and turnaround
WIKA lists broad pressure-instrument capabilities and two Canadian laboratory locations; Duncan Instruments Canada lists digital and analogue gauge calibration and a conditional 48-hour service.[1][4] Compare those offers against the range, uncertainty and certificate evidence your job requires—not simply whether the gauge can be serviced.

WIKA
- Published capabilities: WIKA advertises absolute- and gauge-pressure capability from −1 to 1,000 bar, lists dial gauges, transmitters and differential instruments, and accepts instruments from other manufacturers.[1] Confirm coverage for the particular instrument and pressure mode rather than assuming the entire equipment list is covered across that range.
- Accreditation arrangements: WIKA describes its Edmonton laboratory as ISO 9001 registered and its Oakville laboratory as ISO/IEC 17025 accredited.[1] Request the applicable pressure scope for the location handling the work.[3]
- Turnaround: WIKA advertises fast turnaround, efficient scheduling and an express-service enquiry; the cited page gives no numerical completion time.[1]
Duncan Instruments Canada
- Published capabilities: Duncan advertises calibration of digital and analogue pressure gauges used in gas, air and fluid systems.[4] Ask Duncan to confirm numerical pressure limits and coverage for the required gauge, absolute or differential mode.
- Accreditation arrangements: Duncan describes an ISO 9001:2015-registered laboratory and offers ISO/IEC 17025-accredited calibration in collaboration with an accredited testing laboratory.[4] Ask it to identify the collaborator, applicable scope and certificate issuer.[3][4]
- Turnaround: Duncan states, “We can calibrate your instrument in 48 hours if required.”[4] Obtain confirmation this conditional offer applies to the requested work and deadline.
Make the choice conditional on written acceptance of the calibration specification and certificate requirements.[3]
Check accreditation against the exact work
JM Test cautions that accreditation may cover some instrument types, ranges or methods without covering every service a laboratory offers.[3] Check the proposed work in this order:
- Identify the performing laboratory and location. Establish which facility will handle the gauge before checking its scope.[3]
- Obtain its current accredited scope. Match the instrument category, pressure mode, range, method and uncertainty to the requested calibration.[3]
- Confirm the certificate issuer. Ask which laboratory will issue it, which accreditation body will appear and whether the requested work falls within its accredited scope.[3]
For WIKA, establish which location will perform the work; for Duncan, identify the collaborating accredited laboratory.[1][4] Ask both providers to confirm job-specific accredited coverage.[1][4]
Check traceability separately. JM Test describes it as a documented, unbroken chain to a recognized reference, with each calibration contributing uncertainty; traceability wording alone does not establish accredited coverage.[3]
Preserve as-found results before adjustment
Calibration compares and documents an instrument’s performance; adjustment changes it to bring readings closer to the correct value.[3] As-found results document the condition before adjustment or repair; as-left results document the condition afterwards.[3]
JM Test explains that as-found readings can reveal drift or an out-of-tolerance condition and support decisions about reviewing earlier measurements, changing calibration intervals or restricting instrument use.[3] Preserve that evidence before authorising changes.
Include this proposed condition in the purchase order:
Record and retain as-found calibration results before any adjustment or repair. If the gauge is out of tolerance, notify us and stop work pending written authorisation. After authorised adjustment or repair, record as-left results and provide both datasets with the calibration certificate.
WIKA offers calibration and adjustment, while Duncan advertises calibration and repair.[1][4] Ask each provider to preserve both datasets and require advance authorisation. Obtain written acceptance of the instruction before shipment.
Specify the gauge, pressure mode and working range
JM Test recommends identifying the instrument and confirming the required range before calibration.[3] Put these details in the service request:
- Identity: manufacturer, model, serial number and customer asset ID.[3]
- Type and range: gauge type, indicated units, full instrument range and required calibration range.[3]
- Pressure mode: gauge, absolute or differential pressure; for differential work, request confirmation of any required static-pressure condition.[1]
WIKA lists differential gauges and differential calibration under static pressure alongside its absolute- and gauge-pressure capabilities.[1] Ask either laboratory to confirm the requested mode and range for the exact instrument.
Add the process medium, wetted materials and connection details as compatibility fields. Manogauge’s product-specification guidance supports identifying the precise model, range, wetted materials and connection; it does not establish either laboratory’s ability to accommodate that configuration.[2] Request written acceptance of any calibration-medium restrictions.
Agree points, pressure directions and hysteresis reporting
Send the required pressure points and units, or ask the laboratory to propose a point schedule for approval before shipment. Include instructions for:
- Pressure direction: whether increasing and decreasing pressure runs are required.
- Repeat runs and dwell time: whether readings must be repeated and how long pressure should be held before recording.
- Zero-point treatment: whether and when zero readings should be recorded.
- Hysteresis reporting: the required data, calculation and acceptance criterion.
Ask WIKA and Duncan to agree the point count, percentage spacing, zero treatment, pressure sequence, repeat runs, dwell time and hysteresis limit for the job.[1][4]
Where pressure direction matters to the assessment, request separately identifiable increasing and decreasing readings in the returned dataset. Include this in the agreed reporting requirements rather than relying on a pass/fail summary.
Set uncertainty, tolerance and the decision rule
Calibration can document performance without establishing suitability for the intended tolerance.[3] Keep three requirements separate:
- Tolerance or governing specification: identify the limits against which results should be assessed.[3]
- Measurement uncertainty: request achievable uncertainty across the specified range and points, then assess whether it supports the intended use.[3]
- Conformity statement: state whether a pass/fail conclusion is required rather than measurement results alone.[3]
WIKA advertises instrument accuracies as precise as 0.008%; this is an instrument-accuracy claim, not a promised calibration uncertainty for the submitted gauge.[1] Ask WIKA and Duncan for job-specific uncertainty values, conformity decision rules and any guard bands.[1][4]
Agree the decision rule in writing: ask how measurement uncertainty will affect pass/fail decisions and whether a guard band will be used.[3] JM Test describes a 4:1 Test Accuracy Ratio as a rule of thumb, not a universal requirement; the appropriate relationship depends on uncertainty, tolerance, decision rules and the applicable quality system.[3]
Keep product approvals separate: a product Certificate of Conformance or approval is distinct from calibration results documenting measured performance and uncertainty.[2][3]
Confirm certificate content, delivery and commercial terms
Request a sample certificate and confirmation that the agreed dataset will appear on the certificate or an identified attachment. Check for:
- Identity: manufacturer, model, serial number, asset ID and calibration date.[3]
- Results: agreed readings, as-found data and applicable as-left data.[3]
- Assessment: tolerance or specification, requested conformity statement and measurement uncertainty.[3]
- Supporting details: reference standards, traceability, method, relevant environmental conditions, laboratory identification and applicable accreditation details.[3]
Request dated commitments for receipt, calibration completion, certificate release and return dispatch. Duncan’s conditional 48-hour calibration claim and WIKA’s express-service language do not establish shipping-inclusive delivery dates.[1][4]
Ask WIKA and Duncan to confirm prices and cancellation terms in writing.[1][4] Request an itemised quotation covering accredited work, adjustment, repair, shipping and express service, together with cancellation conditions and approval requirements for additional charges.
If a next-due date will appear, agree its basis rather than defaulting to annual recalibration: JM Test says interval selection depends on use, history, risk and applicable requirements.[3]
Send the same specification to both laboratories
Use one request to compare written commitments and exceptions. The following template combines the instrument details, certificate requirements and proposed adjustment controls.
Please quote against this specification and identify any exceptions:
- Instrument: [manufacturer, model, serial number, asset ID].
- Mode and range: [gauge/absolute/differential; full and working ranges; units; static-pressure condition if applicable].
- Compatibility: [service medium, wetted materials, connection and calibration-medium restrictions].
- Points and directions: [test points; increasing/decreasing runs; repeat runs; dwell time; zero treatment].
- Hysteresis: [required data, calculation and acceptance limit].
- Assessment: [tolerance or governing specification; required uncertainty; pass/fail requirement; proposed decision rule and any guard band].
- Accreditation: [required scope coverage; performing location, accreditation body and certificate issuer].
- Adjustment control: Record as-found results before adjustment or repair. Obtain written authorisation before proceeding with either.
- Certificate: [as-found and applicable as-left readings; uncertainty, method, standards, traceability and environmental conditions].
- Delivery and quotation: [certificate and instrument-return deadlines; itemised total, shipping, optional work and commercial terms].
WIKA’s published instrument breadth and Duncan’s conditional 48-hour option provide different starting points for comparison.[1][4] Choose on confirmed scope, uncertainty and evidence requirements—not the broadest range or fastest headline. Before shipping, obtain acceptance of the specification and resolve any exceptions.
References