Choose a laboratory sieve shaker only after confirming complete-stack fit, electrical compatibility, sample suitability, method-compatible controls, and the run evidence your lab needs.[1][2] Separate required conditions from optional equipment before comparing machines:

Advertisement
Close-up of various spice canisters in a kitchen setting in Lyon, France.
  • Required fit and controls: Confirm the exact configured stack, electrical supply, and motion and timing controls against the lab’s method.[1][2]
  • Required selection evidence: Define the representative sample and require a demonstration record the lab can assess.
  • Conditional requirements: For wet operation, confirm water, drainage or recycling, and filter arrangements.[2] If reporting or export is required, demonstrate that function separately from setting recall.[1]
  • Optional equipment: Treat observation accessories, spares and software as extras only when the method and documentation needs do not require them.
Advertisement

List the actual sieve frames, pan and lid

Start with the complete assembly, not nominal sieve diameter alone. Frame height, sieve count and accessories affect the configuration a machine must accommodate.[1][2] Use this requirements sheet for each candidate:

  • Test method: Designation and version; required sieve opening sequence, top to bottom.
  • Sieve frames: Diameter, full-height or half-height designation for each frame, and sieve count.
  • Component identifiers: Model, serial number or lab inventory identifier, where available, for each sieve and accessory.
  • Required stack components: Pan, lid or cover, gaskets and other method-required components.
  • Optional observation accessories: Spacer or observation cylinder, intended purpose and position in the assembly.
  • Complete assembled height: Measured with every component intended for the test.

Record how accessories change the assembly. Gilson’s SS-10 acrylic spacer replaces one sieve for observation or to maintain stack height; it does not add a sieve position.[1] Wet-Vac includes a transparent observation cylinder that nests within the stack.[2]

Advertisement

Verify configured capacity and resolve the Wet-Vac conflict

Gilson specifies that the SS-10 accepts eight full-height or sixteen half-height 8-inch sieves plus a pan; 200 mm sieves can also be used. It uses adjustable stack clamps to secure the sieves.[1] Check frame diameter, frame height, sieve count and usable assembled-stack height separately.

Wet-Vac’s capacity figures conflict: Gilson’s WV-2 and WV-3 model table lists eight half-height 8-inch sieves, while its generic specification lists sixteen half-height 8-inch/200 mm sieves.[2] Do not select the larger figure or assume accessories explain the difference.

Obtain written, model-specific confirmation from Gilson that the complete measured assembly—including the pan, lid, gaskets, spacer or observation cylinder—fits and is securely clamped.[1][2] Keep Wet-Vac capacity unresolved until Gilson reconciles the conflicting counts for the exact model and configuration.

Assign each candidate a status:

  • Confirmed: Written, model-specific confirmation covers the complete assembly and securing arrangement.
  • Unresolved: A capacity conflict, missing dimension or configuration question remains.
  • Incompatible: The documented model limit excludes the required assembly.

Leave missing dimensional limits blank rather than estimating them.

Advertisement

Specify the sample and method—not a size category

A stack that fits still needs evidence of suitability for the material. Add these lab-defined fields before accepting a sample-size description:

  • Material and condition: Identify the material and its condition when loaded.
  • Representative sample amount: Record the mass or volume the lab needs to test.
  • Procedure: Identify the method and version, required openings, and dry or wet operation.

Gilson describes the SS-10 as processing material 2 mm and finer, depending on material type. Use that range as an initial screen. Ask for the maximum or recommended sample amount for your exact stack and have the supplier demonstrate your material and method. [1][2]

Gilson lists ASTM D6913 under Wet-Vac’s “Meets Standard(s).”[2] Have the lab confirm the configuration and procedure needed for its required method rather than treating that listing as sufficient evidence of conformance. Keep representative-sample criteria, project grading limits and test acceptance criteria lab-defined.

Advertisement

Compare controls separately from recorded results

Compare each machine’s motion and timing controls with the specified method before assessing its documentation functions.

  • SS-10 controls: Gilson lists ten vibration-power levels, adjustable total test time, active-vibration time and timed pause intervals, with a maximum time of 99 minutes 59 seconds. Up to 99 operating profiles can be stored in nonvolatile memory.[1]
  • SS-10 motion: Gilson describes multidirectional vibratory agitation, with pause/resume behavior mimicking tapping and producing a three-dimensional effect.[1] This manufacturer description does not establish equivalence to the lab’s tapping procedure.
  • Wet-Vac controls: Vibration, vacuum and water systems are adjustable. Its timer tracks total operating time up to 99 minutes, including interruptions.[2] Compare that behavior with the method’s definition of test duration.

Keep three questions separate:

  • Can settings be recalled? The SS-10 stores operating profiles. Use the saved settings as part of the run record and check reproducibility with the representative sample. [1]
  • Is a run record available? The cited SS-10 page does not describe an exported report, audit trail, result record or demonstration protocol.[1]
  • Are results reproducible? Assess the representative-sample demonstration against the lab’s criteria rather than treating stored profiles as proof.
Advertisement

Check included equipment, utilities and configuration costs

Organize the configuration into included equipment, method-required additions and optional extras. Gilson lists only the shaker as included with the SS-10; test sieves and the clear acrylic spacer are accessories.[1] Wet-Vac includes the shaker, hand-spray nozzle and hose, and clear acrylic observation cylinder.[2]

Wet-Vac supports dry, wet and dry-to-wet operation across a published No. 16–635 opening range, or 1.18 mm–20 µm.[2] That opening range is separate from the sample-load requirement. For wet operation, document:

  • Water arrangement: A 7-gallon recycling tank or tap-and-drain once-through operation.[2]
  • Filter arrangement: Appropriate filter media and provision to monitor or replace filter paper beneath the stack during a cycle.[2]
  • Electrical compatibility: Wet-Vac models WV-1, WV-2 and WV-3 are rated 115 V, 60 Hz, 16 A; WV-1F, WV-2F and WV-3F are rated 230 V, 50 Hz, 16 A.[2] Verify the selected model against the lab supply. For the SS-10, request written confirmation of electrical compatibility for the exact model.[1]
  • Accessory identification: An itemized configuration with names and model identifiers, rather than assumptions about similarly labeled parts.

Request an itemised price for the complete configuration. For the SS-10, list the required sieves, pan, cover, seals and any observation spacer separately. For Wet-Vac, include the required filters and support equipment for the selected water and drainage arrangement. [1][2]

Advertisement

Agree on a representative demonstration and run record

Request a demonstration using the lab’s actual sieves, complete configured stack, representative material and specified method—not a substitute setup chosen for convenience.

Have the lab define the repeat runs, tolerances and acceptance criteria before the demonstration. Ask the supplier to show which results can be exported and which the operator must record.

Use this proposed demonstration record:

  • Equipment and stack: Machine model and serial number; sieve identifiers and opening sequence; frame diameters and heights; pan, lid, gaskets, spacers, observation cylinder and other fitted accessories.
  • Sample and method: Sample identifier, material description, condition and amount; method designation and version; operator and test date.
  • Operating record: Actual settings, selected profile where applicable, total elapsed and active operating time, programmed pauses, unplanned interruptions and setting changes.
  • Wet-operation record: Water source, recycling or once-through arrangement, water and vacuum settings, filter media, and filter checks or replacements. These fields reflect Wet-Vac’s adjustable systems and filter-monitoring provisions.[2]
  • Results and evidence: Method-required results, observations, deviations, repeat-run comparisons, and identifiers for retained reports and data files.

Request an example of every claimed report or export. Identify whether it is manually completed, equipment-generated or software-generated, and specify which deliverables must accompany the demonstration.

Request the model-specific documentation, electrical specifications and a sample of any report format required for your records. Keep those materials with the demonstration results.

Advertisement

Select only after the required checks are resolved

Use the completed requirements sheet and demonstration record to make the decision:

  • Fit: The exact model accepts the complete configured stack, and any conflicting capacity figures are resolved.
  • Operation: Electrical supply, motion, timing and any wet-operation arrangements match the intended use.
  • Sample suitability: The demonstration addresses the lab’s material, sample amount and specified method.
  • Evidence: The retained record supports the lab’s assessment of results and repeat runs.
  • Configuration: Required accessories and documentation functions are included in the comparison, not left among optional extras.

Keep a candidate on hold if stack capacity, sample suitability or reporting remains unresolved. Compare complete configurations—not nominal diameter, sample-size labels or stored-profile counts alone.